BEMPU Hypothermia Alert Device versus Conventional Care
Year
2025
Location
🇧🇩 Dhaka, Bangladesh
Sample Size
98 LBW infants analyzed
Investigators
Azmery Saima, Abdullahel Amaan, Md. Ashraful Islam, Mohtarama Mostari, Afroza Islam Shuma, Ismat Jahan, M. A. Mannan, Mohammod Shahidullah
Institutions
Bangabandhu Sheikh Mujib Medical University (BSMMU), Dhaka · National Gastroliver Institute & Hospital, Dhaka · Dept of Neonatology, BSMMU
Published In
Reproductive, Female and Child Health
Background & Methods
Key Results
| Outcome / Metric | BEMPU Group | Control Group | p-value |
|---|---|---|---|
| Hypothermia detection events (mean) | 6.73 | 2.87 | 0.001 |
| KMC duration (hrs/day) | 7.19 | 3.04 | 0.001 |
| Weight gain (g/day) | 20.96 | 16.17 | 0.007 |
- 1TempWatch detected significantly more hypothermia events (6.73 vs 2.87, p=0.001), confirming the benefit of continuous monitoring over conventional intermittent checks.
- 2KMC duration was more than double in the TempWatch group (7.19 vs 3.04 hrs/day, p=0.001), consistent with findings from India and Togo.
- 3Daily weight gain was significantly higher in the TempWatch group (20.96 vs 16.17 g/day, p=0.007).
Why This Matters
How to Cite
Related Studies
Role of Hypothermia Detecting Device during Transitional Adaptation of Healthy Full-Term Infant
India · 2021
The Role of a Neonatal Hypothermia Alert Device in Promoting Weight Gain in Low-Birth-Weight Infants
India · 2020
Comparison between Temperature Watch and Mercury Thermometer in Monitoring Temperature of Low-Birth-Weight Newborns at Kangaroo Mother Ward, Chandaka Medical College, Larkana
Pakistan · 2022